ASSESSMENT OF GRIP FORCE CONTROL IN PATIENTS WITH MUSCULAR DYSTROPHY

  • Gregorij Kurillo Faculty of Electrical Engineering University of Ljubljana Tržaška 25 1000 Ljubljana Slovenia
  • Tadej Bajd Institute for Rehabilitation Republic of Slovenia Linhartova 51 1000 Ljubljana Slovenia
  • Anton Zupan Institute for Rehabilitation Republic of Slovenia Linhartova 51 1000 Ljubljana Slovenia
Keywords: grasping, grip force, hand functionality, muscular dystrophy, tracking

Abstract

Background. The majority of hand functionality tests are based on qualitative assessment which largely depends on the experience of the therapist. Computer-assisted methods can provide more objective and accurate measurements of the grip force and other parameters related to grasping.

Methods. We analysed the grip force control in 12 patients with muscular dystrophy using the tracking system developed. The system consists of a grip-measuring device with endobjects assessing the force applied in different grips. The device was used as input to a tracking task where the patient applied the grip force according to the visual feedback from the computer screen. Each patient performed two tasks which consisted of tracking a ramp and sinus target.

Results. We analysed the maximal grip force as assessed in the ramp task and the tracking accuracy of the sinus task. The results are compared among five different grips (cylindrical, lateral, palmar, pinch and spherical grip), applied with dominant and non-dominant hand. The results show no significant difference in tracking accuracy between the dominant and non-dominant hand.

Conclusions. The results obtained in tracking the ramp target showed that the method could be used for the assessment of the muscle fatigue, providing quantitative information on muscle capacity. The results of the sinus-tracking task showed that the method can evaluate the grip force control in different types of grips, providing information on hand dexterity, muscle activation patterns or tremor.

Downloads

Download data is not yet available.

References

Napier JR. The prehensile movements of the human hand. J Bone Joint Surg 1956: 38B; 902–13.

MacKenzie CL, Iberall T. Advances in psychology, 104: The grasping hand. Amsterdam: Elsevier Science B.V., 1994.

McPhee S. Functional hand evaluations: a review. Am J Occup Ther 1987; 41: 158–63.

Fugl-Meyer AR, Jääskö L, Leyman I, Olsson S, Steglind S. The post-stroke hemiplegic patient. I. A method for evaluation of physical performance. Scand J Rehab Med 1975; 7: 13–31.

Smidt GL, Rogers MW. Factors contributing to the regulation and clinical assessment of muscular strength. Phys Ther 1982; 62: 1283–90.

Kattel BP, Fredericks TK, Fernandez JE, Lee DC. The effect of upper-extremity posture on maximum grip strength. Int J Ind Ergon 1996; 18: 423–9.

Innes E. Handgrip strength testing: A review of the literature. Aust Occup Ther J 1999; 46: 120–40.

Jones L. Dexterous hands: Human, prosthetic, and robotic. Presence 1997; 6: 29–57.

Jones RD. Measurement of sensory-motor control performance capacities: Tracking tasks. In: Bronzino JD ed. The biomedical engineering handbook 2nd ed. Vol. II. Boca Raton: CRC Press, 2000.

Blank R, Heizer W, von Voss H. Development of externally guided grip force modulation in man. Neurosci Lett 2000; 286: 187–90.

Vaillancourt DE, Slifkin AB, Newell KM. Visual control of isometric forces in Parkinson’s disease. Neuropsychologia 2001; 39: 1410–18.

Kriz G, Hermsdörfer J, Marquardt C, Mai N. Feedback-based training of grip force control in patients with brain damage. Arch Phys Med Rehabil 1995; 76: 653–9.

Kurillo G, Bajd T, Mihelj M. Force tracking in two-oppositional grips. In: Hutten H, Krösl P eds. Proceedings of the 2nd European Medical and Biological Engineering Conference – EMBEC’02. Vienna: Druckerei Agath, 2002. II/1712–3.

Chadwick EKJ, Nicol AC. A novel force transducer for the measurement of grip force. J Biomech 2001; 34: 125–8.

Fowler NK, Nicol AC. A force transducer to measure individual finger loads during activities of daily living. J Biomech 1999; 32: 721–5.

Memberg WD, Crago PE. Instrumented objects for quantitative evaluation of hand grasp. J Rehab Res Dev 1997; 34: 82–90.

Kurillo G, Bajd T, Kamnik R. Static analysis of nippers pinch. Neuromodulation 2003; 6 (in Press).

Smith SS. Measurement of neuromuscular performance capacities. In: Bronzino JD ed. The biomedical engineering handbook 2nd ed. Vol. II. Boca Raton: CRC Press, 2000.

How to Cite
1.
Kurillo G, Bajd T, Zupan A. ASSESSMENT OF GRIP FORCE CONTROL IN PATIENTS WITH MUSCULAR DYSTROPHY. TEST ZdravVestn [Internet]. 1 [cited 5Aug.2024];73. Available from: http://vestnik-dev.szd.si/index.php/ZdravVest/article/view/2404
Section
Original article